React Health updates Luna G3 BPAP correction, reports two serious adverse reports

React Health issued a November 17 update on its Luna G3 BPAP device correction, reporting 15 complaints and two serious adverse reports linked to inaccurate pressure-data reporting when Targeted VT is enabled.

The company notice posted by the FDA concerns specific Model LG3800-30VT devices that fail to record or transmit actual pressure statistics when Targeted VT is enabled. Inaccurate historical reports could prompt inappropriate settings adjustments and delayed therapy. React Health has developed firmware G3-2.00.80 and is coordinating installation through equipment providers; the update does not correct previously recorded data.

Official · 2 sources

Evidence

  1. OfficialReact Health published a correction update on November 17, 2026.
    “Update Published: November 17, 2026”

    Source:FDA

  2. OfficialReact Health published the correction update November 17.
    “Update Published: November 17, 2026”

    Source:FDA